In our #100Messmittel series, we’re showcasing 100 measuring and testing devices that we calibrate every day at QMK. So far, we’ve covered the vernier caliper and the limit plug. Today, we’re taking it a step further in terms of precision.
🔥 With this measuring instrument, the heat from your own hand can skew the result.
No joke: A steel part that you hold in your hand for too long expands. With a length of 100 mm and a temperature difference of just a few degrees, we’re talking about several micrometers. That might not sound like much—but it’s exactly the range in which this device operates.
Measuring instrument No. 3 in our series: the micrometer.
What is a micrometer?
Where the vernier caliper stops at tenths of a millimeter, the micrometer begins. Thanks to a precise measuring spindle with a defined thread, it provides resolution in the micrometer range—which is why almost everyone calls it a “micrometer.” It is the instrument of choice for critical dimensions: shafts, fits, and sealing surfaces.
The principle is simple and ingenious: One revolution of the measuring spindle moves it by the thread pitch—0.5 mm on most devices. The scale drum divides this revolution into 50 divisions—equaling 0.01 mm per division. Digital micrometers usually have a resolution of as fine as 0.001 mm.
Overview of the Design
- Bracket: supports the anvil and measuring spindle. It is often fitted with insulating plates—precisely because of the heat from the hands.
- Anvil and measuring spindle: feature the measuring surfaces, usually tipped with carbide to minimize wear.
- Scale sleeve and scale drum or digital display.
- Ratchet or spring-loaded ratchet: limits the measuring force so that each measurement is performed with the same force.
- Locking mechanism: secures the spindle, e.g., for serial measurements.
Clamp micrometers are available in measuring ranges of 25 mm each (0–25 mm, 25–50 mm, 50–75 mm, etc.). For larger ranges, an adjustment gauge is included to set the zero point or starting value.
Precision Comes at a Price: The Factors That Affect It
The measuring force, the cleanliness of the measuring surfaces, the temperature of the component, the temperature of the measuring instrument—at this level of resolution, everything becomes a influencing factor that you could safely ignore when using a vernier caliper.
- Temperature: Steel expands by about 11.5 µm per meter per Kelvin. A 100-mm part that warms up by 3 °C in your hand is therefore already a good 3 µm longer. The reference temperature for length measurements is 20 °C.
- Measuring force: A device tightened too tightly will systematically measure too low. Therefore, always measure using the ratchet; never “push down” on the drum.
- Cleanliness: Even a thin film of oil or a single speck of dust on the measuring surfaces can skew the result by several micrometers.
- Handling: Hold the micrometer by the insulated handle and allow it to acclimate for a sufficient amount of time before taking a measurement.
How do we calibrate a micrometer?
We check the display accuracy across the entire measuring range against traceable reference gauges. To do this, we check the flatness and parallelism of the measuring surfaces—we examine these using a magnifying glass, where deviations become visible as interference fringes. And we check the measuring force, because a device that is tightened too tightly will systematically measure too low.
The procedure in detail:
- Cleaning and Temperature Control: The device and standards are brought to 20 °C in the climate-controlled laboratory.
- Visual and functional inspection: Smooth operation of the spindle, function of the ratchet and locking mechanism, condition of the measuring surfaces, zero point.
- Flatness of the measuring surfaces: An optical flat glass plate is placed on the measuring surface. The resulting interference fringes indicate deviations from the plane—each fringe corresponds to only about 0.3 µm.
- Parallelism of the measuring surfaces: Using plane-parallel glass plates of varying thicknesses, so that the spindle is tested in different rotational positions.
- Display accuracy: Using gauge blocks at test points distributed across the measuring range and covering various spindle positions—this also reveals periodic pitch errors in the measuring spindle.
- Measured force: Check whether the ratchet triggers within the permissible range.
- Setpoint: If available, the corresponding setpoint is also calibrated.
- Documentation: Measurement deviations with measurement uncertainty are recorded in the calibration certificate; a statement of conformity is provided upon request.
The basis for this is the product standard DIN EN ISO 3611 for micrometers. You can find more information on traceability and measurement uncertainty in the QMK Calibration Glossary.
Where is the micrometer used in practice?
Anywhere where tolerances are tighter than a few hundredths: shaft diameters, bearing seats, sealing surfaces. In quality assurance, it’s often the final word before approval—whatever it says goes, or it doesn’t.
A device capable of measuring in micrometers must itself be accurate down to the micrometer level. Otherwise, you’ll be measuring very precisely—but incorrectly. At QMK, we’ve already calibrated over 12,500 micrometers.
And because even the best measuring instrument is only as good as the hand that operates it: In our training courses and seminars, we show you what really matters when it comes to measuring.
Frequently Asked Questions About Calibrating Micrometers
How often should a micrometer be calibrated?
There is no legally mandated interval. It is common to start with a 12-month interval, which is then adjusted based on usage, the environment, and previous calibration results. Equipment used daily in manufacturing typically requires shorter intervals.
Is it sufficient to check the zero point yourself?
Checking the zero point before measuring is important, but it only provides information about a single point. This method does not detect lead errors in the spindle, measuring surfaces that are no longer flat or parallel, or an incorrect measuring force.
Does the reference dimension need to be calibrated as well?
Yes. For micrometers with a range of 25 mm or more, the starting value is set using the calibration gauge. If the calibration gauge is incorrect, every measurement will be off by the same amount.
How do I send my micrometers to QMK?
By mail, in person, or via our pickup and delivery service.
Have your micrometers calibrated at QMK
As a DAkkS-accredited calibration laboratory (D-K-20571-01-00) based in Offenburg, we calibrate micrometers of all types—analog and digital, with and without a set gauge. Feel free to contact us.