Calibrating Pressure Gauges: Hysteresis, Accuracy Class, and Pressure Units | #100Measuring Instruments No. 6

In our #100Messmittel series, we’re showcasing 100 measuring and testing devices that we calibrate every day at QMK. After length, electrical parameters, and torque, today we’re focusing on the fourth measurement parameter: pressure.

🔴 A pressure gauge always shows you a pressure reading. The only question is whether it’s the correct one.

The needle stands firmly at “6 bar”—and no one at the plant doubts it. But that’s exactly the trap: a pressure gauge gradually loses its accuracy. You can’t tell just by looking at it.

Measuring instrument No. 6 in our series: the pressure gauge.

Calibration on the pressure test bench using a digital reference pressure gauge in the QMK laboratory
On the pressure test bench, the test specimen is compared to a feedback-controlled reference device.

What is a pressure gauge?

The classic pressure gauge uses a Bourdon tube, also known as a Bourdon spring: A bent tube stretches slightly under pressure, and this movement drives the pointer via a pointer mechanism. Simple and robust—but the tube spring ages, develops hysteresis, and eventually displays a different reading when pressure rises than when it falls.

Bourdon tube pressure gauge in accordance with EN 837-1, with a scale up to 100 bar and accuracy class 1.0
Bourdon tube pressure gauges in accordance with EN 837-1—the accuracy class is marked directly on the dial (“CL 1.0”).

What does the accuracy class mean?

The dial of many pressure gauges displays a class, such as “CL 1.0.” This indicates the permissible deviation as a percentage of the measuring range. For a Class 1.0 pressure gauge with a measuring range of 0 to 100 bar, the reading may therefore deviate by up to 1 bar at any point on the scale—at an operating pressure of 6 bar, that’s already about 17% of the measured value. That’s why it’s a good idea to choose a measurement range so that the operating pressure isn’t right at the bottom of the scale.

How do we calibrate a pressure gauge?

Compared to a traceable pressure standard, across multiple test points—and deliberately in both directions: rising and falling. This is exactly how the hysteresis becomes apparent, which would be overlooked in a single measurement. The calibration certificate lists the actual deviation at each point, not just at the end of the scale.

An overview of the procedure:

  1. Visual and functional inspection: Condition of the case, dial, connection threads, and pointer; zero point when the gauge is unpressurized.
  2. Pre-loading: Before measurement, the pressure gauge is loaded multiple times up to the upper limit of the measurement range.
  3. Ascending measurement series: Several test points across the entire measurement range are reached sequentially and compared with the reference value.
  4. Descending measurement series: The same points in reverse order. The difference between the two series indicates the hysteresis.
  5. Evaluation and Documentation: Deviation, hysteresis, and measurement uncertainty for each test point; upon request, with an assessment against the accuracy class.

With mechanical pressure gauges, gently tap the housing before taking a reading. This reveals whether friction in the pointer mechanism is causing the pointer to “stick.” You can find more information on traceability and measurement uncertainty in the QMK Calibration Glossary.

Analog pressure gauges and digital pressure transducers in a transport case prior to calibration
Analog or digital: Every pressure gauge is tested against a traceable standard.

Where are pressure gauges used in practice?

Anywhere pressure plays a role: on compressors and hydraulic systems, in process engineering, on test benches, and in leak testing. A pressure gauge that displays incorrect readings can render an entire system acceptance test or safety inspection invalid.

Fun fact: bar, psi, kPa, mmHg—the variety of units used for pressure

Hardly any other physical quantity has as many units as pressure. Every day, we juggle between bar, psi, kPa, and even mmHg—depending on the customer, industry, and country of origin of the device. And that’s no small matter: 1 bar is roughly 14.5 psi. If you mix the two up, you’re not just a little off—you’re off by a factor of 14. Keeping track of units accurately is part of our craft.

Unit Equivalent to 1 bar Typical use
bar 1 bar Industry and Technology in Europe
kPa 100 kPa SI units, standards, and science
psi approx. 14.5 psi Devices from the U.S., tire pressure
mmHg approx. 750 mmHg Medicine, vacuum technology
Top view of a box containing analog pressure gauges and digital pressure measuring devices for calibration
We have already calibrated 2,258 pressure measuring devices—ranging from analog pressure gauges to digital measuring instruments.

A reading that looks convincing is by no means an accurate value. Only those who calibrate know the difference.

By the way: Digital pressure transmitters that send their signal directly to the plant control system are a topic all their own—we’ll take a closer look at them in a later post.

Frequently Asked Questions About Calibrating Pressure Gauges

How often should a pressure gauge be calibrated?

There’s no set interval. Twelve months is a common starting point. In cases of pressure spikes, vibrations, pulsating pressure, or aggressive media, a shorter interval may be advisable—as well as after an overload or if the pointer no longer reads zero when the system is depressurized.

What is hysteresis in a pressure gauge?

Hysteresis is the difference in the reading at the same pressure, depending on whether the pressure is approached from below (rising) or from above (falling). It results from the material properties of the spring and friction in the pointer mechanism—and is only noticeable when measurements are taken in both directions.

Do you also calibrate digital pressure gauges?

Yes—from analog pressure gauges to digital handheld meters to pressure transducers. We’ll be dedicating a separate article to the latter in this series.

How do I get my pressure gauges to QMK?

By mail, in person, or via our pickup and delivery service.

Have your pressure gauges calibrated at QMK

As a DAkkS-accredited calibration laboratory (D-K-20571-01-00) based in Offenburg, we calibrate pressure gauges and pressure measuring instruments—both analog and digital—using ascending and descending measurement ranges. Please feel free to contact us.